Analytical Sciences


Abstract − Analytical Sciences, 31(11), 1177 (2015).

Separation of Metal Complexes with Counter Ions by Tube Radial Distribution Chromatography Using a Ternary Solvent Containing 8-quinolinol
Yuji KAWAI,* Satoshi FUJINAGA,* Masahiko HASHIMOTO,* and Kazuhiko TSUKAGOSHI*,**
*Department of Chemical Engineering and Materials Science, Faculty of Science and Engineering, Doshisha University, Kyotanabe, Kyoto 610-0321, Japan
**Tube Radial Distribution Phenomenon Research Center, Doshisha University, Kyotanabe, Kyoto 610-0321, Japan
An open-tubular capillary chromatography system (tube radial distribution chromatography, TRDC) was developed using a ternary solvent (water-acetonitrile-ethyl acetate; volume ratio, 3:8:4) containing 10 mmol L−1 8-quinolinol for the separation of nitrate, chloride, and sulfate compounds of Ni(II), Al(III), and Fe(III). When a mixed solution of the Ni(II) compounds was injected into an untreated fused-silica capillary tube (90 cm × 75 μm i.d.) with a ternary solvent flow rate of 0.8 μL min−1, the compounds were eluted in the following order: [Ni(II)-(8-quinolinol)3] complex, [Ni(II)-(8-quinolinol)]-nitrate ion interaction complex, [Ni(II)-(8-quinolinol)]-chloride ion interaction complex, and [Ni(II)-(8-quinolinol)]-sulfate ion interaction complex. The elution of mixtures of the Al(III) and Fe(III) compounds showed similar trends.