In order to compare and evaluate the atomization process occurring in several types of graphite furnaces for atomic absorption spectrometry, the authors estimated temporal variations in the gas temperature by using a two-line method under the assumption of a Bolzmann distribution. The atomization furnaces employed were a graphite tube, a graphite tube coated with pyrolitic carbon, a graphite tube with a platform and a graphite cup. Differences in the temporal variation in the gas temperature among these graphite furnaces were observed.