Abstract − Analytical Sciences, 24(9), 1099 (2008).

A Coulometric Analysis Method and an Ion-exclusion Chromatographic Method for the Determination of Antimony(V) in Large Excess of Antimony(III)
Akiharu HIOKI
National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 3-9, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan
A coulometric analysis method and an ion-exclusion chromatographic method were developed for the determination of antimony(V) in a large excess of antimony(III). Antimony(V) reacted with potassium iodide in a high concentration hydrochloric acid; the liberated iodine was determined by the standard-addition method using coulometrically generated iodine. Using a Dionex ICE-AS1 ion-exclusion column, antimony(V) was eluted with 40 mmol/L sulfuric acid; on the other hand, antimony(III) was strongly retained on the column. The content, expressed as the amount ratio of antimony(V) to antimony(III), was 0.035% in a 10 g/kg antimony(III) solution prepared from an antimony(III) oxide reagent by the coulometric analysis method and 0.036% in a 1 g/kg antimony(III) solution prepared from the same antimony(III) oxide by the ion-exclusion chromatographic method. The results of both methods were in good agreement with each other. The detection limit of antimony(V) in antimony(III) oxide by the former method was 0.004% of antimony(III), and that by the latter method was 0.002% of antimony(III).