Analytical Sciences


Abstract − Analytical Sciences, 24(5), 601 (2008).

Direct Introduction of Slurry Samples in Multi-syringe Flow Injection Analysis: Determination of Potassium in Plant Samples
M. Inês G. S. ALMEIDA,* Marcela A. SEGUNDO,** José L. F. C. LIMA,** and António O. S. S. RANGEL*
*Escola Superior de Biotecnologia, Universidade Católica Portuguesa, Rua Dr. António Bernardino de Almeida, 4200-072 Porto, Portugal
**REQUIMTE, Serviço de Química-Física, Faculdade de Farmácia, Universidade do Porto, Rua Aníbal Cunha, 164, 4050-047 Porto, Portugal
The present work explores the slurry sampling approach for automatic, flow-based plant analysis. For this purpose, pinch valves were introduced into a multi-syringe flow injection analysis manifold to provide the repeatable aspiration of a few microliters of plant suspension before the material was further processed through the flow system. For validation of the proposed approach, the determination of potassium by flame emission spectrometry was implemented. Several parameters were studied: the concentration of plant particles in the sample suspension and the utilization of matrix modifiers. Microwave digestion was also implemented; no significant difference was found when certified reference material was analyzed with or without the in-line digestion step. The system was successfully applied to 13 samples within a concentration range of 2.5 to 100 mg g-1. A determination frequency of 28 h-1 was achieved and the precision was better than 4.0% (n = 12).