Analytical Sciences


Abstract − Analytical Sciences, 22(1), 91 (2006).

Electrochemical Determination of Minoxidil in Different Pharmaceutical Formulations by Flow Injection Analysis
Valeria PFAFFEN and Patricia I. ORTIZ
INFIQC, Departamento de Fisicoquímica, Facultad de Ciencias Químicas, Universidad Nacional de Córdoba, Ciudad Universitaria, 5000 Córdoba, Argentina
The electrochemical behavior and amperometric-FIA quantification of minoxidil at a glassy carbon electrode is described. The procedure is based on electrochemical oxidation at 0.800 V (vs. Ag/AgCl/NaCl(3 M) in a phosphate buffer solution. Minoxidil was determined over the range 1 × 10-7 - 1 × 10-4 M. Different analytical parameters and electrode stability were analyzed to obtain the best electrode performance. The optimal conditions were: working potentials, 0.800 V; flow rate, 0.74 mL min-1; and solution pH 7.0. This system allowed a sampling rate of 120 samples per hour without any pretreatment. The proposed method was used for minoxidil quantification in pharmaceutical preparations with satisfactory results. The accuracy of FIA-amperometric method was established by a comparison with the conventional UV determination technique using a paired t-test indicating the absence of systematic errors.