Analytical Sciences


Abstract − Analytical Sciences, 21(5), 545 (2005).

Flow Injection Determination of Formaldehyde by Its Catalytic Effect on the Oxidation of Sulfonazo III by Bromate with Spectrophotometric Detection
Ali Asghar ENSAFI and Ebrahim HONARMAND
College of Chemistry, Isfahan University of Technology, Isfahan 84156, Iran
A simple and sensitive flow injection method with spectrophotometric detection was developed for the determination of formaldehyde. The method is based on the catalytic effect of formaldehyde on the oxidation of sulfonazo III with bromate in acidic media. The decrease in absorbance of the reaction mixture was measured at 566 nm. The calibration graph was linear in the range of 0.005 to 2.80 µg ml-1 formaldehyde at a rate of 38 ± 4 samples h-1. The limit of detection was 4 ng ml-1. The relative standard deviations for ten replicate measurements of 0.20, 0.50 and 1.00 µg ml-1 formaldehyde were 1.3, 0.8 and 0.7%, respectively. The method was applied to the determination of formaldehyde in river water, shampoo and melamine-formaldehyde resin.