Abstract − Analytical Sciences, 17(4), 523 (2001).
A New Approach of Direct Sample-Digestion before Vaporization for Determination of a Metal in Rocks by Inductively Coupled Plasma Atomic Emission Spectrometry
Yasuaki OKAMOTO, Reiko KIKKAWA, and Terufumi FUJIWARA
Department of Chemistry, Graduate School of Science, Hiroshima University, Higashihiroshima 739-8526, Japan
A new approach to sample digestion, subsequent vaporization and introduction to an inductively coupled plasma (ICP) atomic emission spectrometer was developed for the direct determination of magnesium. To each small sample cuvette made of tungsten, a ground rock sample was precisely weighed. The cuvette was situated on a tungsten boat furnace. Ammonium fluoride solution was added to the cuvette as a chemical modifier. After the on-furnace digestion has been completed, the analyte, magnesium, in the cuvette was vaporized and introduced into the ICP atomic emission spectrometer. Since the powdered samples were wet-digested in the sample cuvettes prior to vaporization, they could be analyzed by using a calibration curve prepared from aqueous standard solutions. This method was applied to the determination of magnesium in several standard reference materials with satisfactory results.
J-STAGE:
View this article in J-STAGE