Analytical Sciences


Abstract − Analytical Sciences, 14(5), 971 (1998).

High-Performance Liquid Chromatographic Determination of Oxalic Acid in Tea Using Tris(1,10-phenanthroline)- ruthenium(II) Chemiluminescence
Fengwu WU, Zhike HE * , Qingyao LUO and Yun'e ZENG
Department of Chemistry, Wuhan University, Wuhan 430072, P. R. China
The chemiluminescence (CL) of Ru(phen)32+ was applied to HPLC determination of oxalic acid, which was separated by a C18 reverse-phase column with a mobile phase of 0.02 mol/l NH4Ac, mixed with 0.25 mol/l Ru(phen)32+ and 2.0 mmol/l Ce(SO4)2 (0.08 mol/l H2SO4), and then passed through a homemade CL detector. The reaction of Ce(IV) oxidized Ru(phen)32+ and the oxalic acid emitted light. The detection limit was 6.2x10-6 mol/l for oxalic acid at a S/N ratio of 3, the relative standard deviation for 5 replicate injections of 1x10-3 mol/l oxalic acid standard was calculated as 5.6%, and the linear calibration range was 1x10-5 to 4x10-3 mol/l. The method was successfully applied to determination of oxalic acid in tea. (Keywords: Oxalic acid, chemiluminescence, Ru(phen)32+, high-performance liquid chromatography, tea)